TSV302 microcomputer-controlled thermal deformation Vica softening point testing machine

Function summary:

      TSV302 microcomputer-controlled thermal deformation Vica softening point testing machine equipment integration is high,The configuration of the implementation components and sensors make the equipment with high reliability, high temperature resistance, small deformation, sensitive response, accurate data transmission and other characteristics;The use of the equipment with the chiller can double the utilization rate of the equipment,The sample frame is made of low expansion coefficient alloy,Within 200℃, it can meet the standard requirements without correction。The machine is widely used for testing Vica softening temperature and load deformation temperature of thermoplastics and polymer materials。It is especially suitable for the production inspection, development and research of products and materials by pipe manufacturers, product quality checkpoints, construction engineering quality checkpoints, construction companies, scientific research institutions and other units。


Relevant standards:

  • GB/T8802-2001 Thermoplastic pipe, pipe fittings Vica softening temperature test

  • GB/T1633-2000 Thermoplastics Vica softening temperature (VST) determination

  • GB/T1634.1-2019 Plastics - Determination of deformation temperature under load - Part 1: General test methods

  • GB/T1634.2-2019 Plastics - Determination of deformation temperature under load - Part 2: Plastic ebonite and long fibre reinforced composites

  • D1525-00 Stand Test Method for Vicat Softening Temperture of Plastics

  • D648-01 Stand Test Method for Deflection of Plastics Under

  • Flexural Load in the Edgewiae Position


Equipment features:

  • The main sensors and execution components equipped with this machine have professional measurement performance of high quality and high reliability。

  • The thermal deformation temperature and Vica softening temperature can be tested simultaneously。

  • The initial cooling temperature is high, up to 280℃, which greatly improves the equipment utilization rate。

  • The sample frame has very little deformation under heat, and the deflection temperature without correction can reach 200℃。

  • The sample rack has automatic loading and automatic lifting functions, which eliminates manual handling of weights。

  • The fuel tank volume is reduced by more than 50% compared with the traditional equipment, which effectively reduces the user's use cost。


Technical parameters:

Specification and model

TSV302A

TSV302B

TSV302C

Test station

3

4

6

Temperature collection point

3

4

6

Tank volume

9L

9L

14L

Cooling time

≤45 minutes (280 ° C to 25 ° C, use with a chiller。)

Cooling mode

水冷

Load loading

电动

Temperature control range

Ambient temperature ~300℃

Maximum temperature error

±0.5℃

Deformation measuring range

0mm to 10mm

Deformation measurement resolution

0.001mm

Deformation measurement accuracy

0.003mm

Deflection without correction temperature  

20℃-200℃

VST Maximum sample size

50mm×16mm×6.5mm (L×W×H)

HDT maximum sample size

130mm×15mm×13mm (L×W×H)

HDT span

64mm, 100mm, 65.6mm、101.6mm

Vicar load

GA=10N±0.2N;GB=50N±1N

Bending stress

1.80MPa, 0.45MPa, 8.00MPa

Rate of warming

(12 ±1)℃/ 6min [(120±10)℃/h]

(5±0.5)℃/ 6min [(50±5)℃/h]

Heating medium

201 dimethylsilicone oil/Ambient temperature ~300℃User provided

10# transformer oil/Ambient temperature ~200℃User provided

Heating power

4.5kW

电源

380V±10% 50Hz 5kW

重量

115kg

180kg